Nursery Equipment

Kay & Company offers a wide range of world-class, CE marked Nursery Equipment. The range caters to all specialties, including Infant Incubator-Neonatal Intensive Care Incubator.

     1   2   3   4   Next »

INFANT INCUBATOR - Neonatal Intensive Care Incubator

Neonatal Intensive Care Incubator
Model: 100A

Specification:

Working temperature range

Accuracy
Resolution
Accuracy of probe Interchangeability
Need for probe calibration
Temperature Probe
Voltage
Power
Heating element
Temperature display
Message and alarms

: 32° to 38° C (patient mode)
: 30° to 38° C (Air mode)
: +/- 0.2° C
: 0.1° C
: +/- 0.2° C
: Not required
: Thermistor based interchangeable probe
: 220 to 250V at 50 Hz
: 500 Watts
: Encapsulated heater with aluminum fins.
: Bright Numerical LED display
: Alphanumerical display

Performance Specifications:

Temperature rise time
Temperature variability
Temperature overshoot
Bed tilting position

: 20-30 min per 10° C
: 0.2° C
: 0.8° C
: + 8° C maximum continuously variable on both sides

Oxygen Concentration:

Oxygen input
5 litres min-1
10 litres min-1
15 litres min-1
Humidification (Optional)
Range
Display resolution

Average concentration range
: 25 to 45%
: 35 to 60%
: 45 to 70%
: Using controller
: 40-90 + 5%
: + 5%

Audio and visual display alarm conditions:
* Patient temperature (more than + 0.5° C difference) * Control temperature
* High Air temperature (more than + 0.5° C difference) * Circular fan failure
* Probe failure * System failure * Power failure * Heater failure
(Automatic heater cutoff if the temperature inside the incubator exceeds 40° C)

Coating :
Epoxy/powder coated body for scratch and rust prevention

Dimension (maximum)

Incubator canopy
Length
Width
Heigth
Mattress
Casters


: 820 mm
: 470 mm
: 430 mm
: 650 mm
: 125mm diameter, 2 locking and 2 non-locking

Operational environment:


Operating temperature range
Storage temperature range
Operating humidity range

: 20° C to 32° C
: -25° C to 60° C
: zero to 95%

* Confirms IEC-601 safety standard for medical equipment.
* Microprocessor based electronic system that performs periodical self-diagnosis.
* Easy to operate membrance keypad for better user efficiency.
* Easy to clean air circulation path helps to maintain high level of hygiene.
* Mobile with 4 swivel castors.
    Kay & ComapnySend Enquiry

     1   2   3   4   Next »

OTHER PRODUCTS IN THIS CATeGORY
- Neonatal Intensive Care Incubator
- Neonatal Phototherapy Unit
- Neonatal Radiant Warmer
- Open Care Warmer - Resuscitation Unit
- Transport Incubator