Nursery Equipment
Kay & Company offers a wide range of world-class, CE marked Nursery Equipment. The range caters to all specialties, including
Infant Incubator-Neonatal Intensive Care Incubator.
INFANT INCUBATOR - Neonatal Intensive Care Incubator
|
Specification:
Working temperature range
Accuracy
Resolution
Accuracy of probe Interchangeability
Need for probe calibration
Temperature Probe
Voltage
Power
Heating element
Temperature display
Message and alarms |
: 32° to 38° C (patient mode)
: 30° to 38° C (Air mode)
: +/- 0.2° C
: 0.1° C
: +/- 0.2° C
: Not required
: Thermistor based interchangeable probe
: 220 to 250V at 50 Hz
: 500 Watts
: Encapsulated heater with aluminum fins.
: Bright Numerical LED display
: Alphanumerical display |
Performance Specifications:
Temperature rise time
Temperature variability
Temperature overshoot
Bed tilting position | : 20-30 min per 10° C
: 0.2° C
: 0.8° C
: + 8° C maximum continuously variable on both sides |
Oxygen Concentration:
Oxygen input
5 litres min-1
10 litres min-1
15 litres min-1
Humidification (Optional)
Range
Display resolution |
Average concentration range
: 25 to 45%
: 35 to 60%
: 45 to 70%
: Using controller
: 40-90 + 5%
: + 5% |
Audio and visual display alarm conditions:
* Patient temperature (more than + 0.5° C difference)
* Control temperature
* High Air temperature (more than + 0.5° C difference) * Circular
fan failure
* Probe failure * System failure * Power failure * Heater failure
(Automatic heater cutoff if the temperature inside the incubator exceeds
40° C)
Coating : Epoxy/powder coated body for scratch and rust prevention
Dimension (maximum)
Incubator canopy
Length
Width
Heigth
Mattress
Casters |
: 820 mm
: 470 mm
: 430 mm
: 650 mm
: 125mm diameter, 2 locking and 2 non-locking |
Operational environment:
Operating temperature range
Storage temperature range
Operating humidity range |
: 20° C to 32° C
: -25° C to 60° C
: zero to 95% |
* Confirms IEC-601 safety standard for medical equipment.
* Microprocessor based electronic system that performs periodical self-diagnosis.
* Easy to operate membrance keypad for better user efficiency.
* Easy to clean air circulation path helps to maintain high level of hygiene.
* Mobile with 4 swivel castors. |
| |
|
Send Enquiry |
OTHER PRODUCTS IN THIS CATeGORY
other product range
|