Our Digital HB Reader is a device that helps in getting a fair reading of Haemoglobin
(Hb) count digitally.
Digital-Hb-Reader
Introduction And Description
The Digital Haemoglobin (Hb) Reader is designed, keeping in the view to make
the analysis simpler. The cumbersome step of calculation is deleted while using
the digital Hb Reader. The final result of the unknown test will be displayed
directly in conc. (gm/dl). There is no need of calculations. However O.D. Switch
is also given in case of optical density is required to crosscheck the reading.
A micro controller has been incorporated in the system for accuracy and reproducibility
of findings.
The Digital Hb Reader is a compact, elegant, very reliable micro controller
based equipment to produce accurate results. An optical glass filter of 540
± 10 nanometers and square cuvette have been incorporated to give accurate
results. This Hb Reader has been designed to operate in the following two modes
: -
A. STD.MODE [Concentration of 15 gm/dl has been fed]
B. NON STD. MODE [Absorbance of 0.30 (O.D.) has been fed]
In case, the CONC. or ABS. of STD. are different from 15 gm/dl or 0.30 respectively,
the user can feed the required values with the help of switch no. 3 & 4
(given on the KEYPAD on front panel).
Specifications
Wavelength Filter - 540 ± 10 nm optical glass filter
Minimum Volume - 1 ml
Display - 7 segment, Red LED, 3.5 digit
Indicator - 5 mm LED Indicators provided for different modes.
Lamp - 6 V / 10W Tungsten Lamp
Detector - Si photodiode
Input Supply
Mains Operation - 230V ± 10%, 50Hz, AC
Battery Operation - 12V
Dimension - 12" x 11.5" x 6.25"
Weight - 5 kg Approx.
Standard Accessories
Lamp - 6V / 10 W - 2 Nos.
Fuse - 350 mA - 2 Nos.
Cuvette - Square - 5 Nos.
Dust Cover - Plastic - 1 No.
Instruction Manual - Book Type - 1 No.
Battery Lead - 1 Pair
Installation Requirement
(A) Voltage 230V ± 10 AC, 50Hz (or AC stabilizer)
(B) Earthing must be proper.
(C) Room should be dust free.
(D) Vibration should be avoided (electrical or mechanical)
(E) Radiation should be avoided.
(F) Humidity should be less than 60-70%.
Unpack the equipment and plug into three-pin socket having proper earth. Leave
it for 5-10 minutes to warm up before use.